power supply unit |
AC input range: 100–240V, 50–60Hz 1.0–1.5A, Class II 115V, 400Hz 1.5A, Class II (nominal for aircraft use) DC output: 24V 3.75A Typical power consumption: 53W (57VA) Peak power consumption: 104W (108VA) |
Environmental conditions |
Operating temperature: +5°C to +35°C Note: The air flow for breathing produced by this therapy device can be higher than the temperature of the room. Under extreme ambient temperature conditions (40°C) the device remains safe. Operating humidity: 10 to 95% relative humidity, non-condensing Operating altitude: Sea level to 2,591 m; air pressure range 1013 hPa to 738 hPa Storage and transport temperature: -20°C to +60°C Storage and transport humidity: 5 to 95% relative humidity, non-condensing |
Electromagnetic compatibility |
The AirStart 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2:2007, for residential, commercial and light industry environments. It is recommended that mobile communication devices are kept at least 1 m away from the device. |
Sensors |
Pressure sensor: Internally located at device outlet, analogue gauge pressure type, -5 to +45 cm H2O (-5 to +45 hPa) Flow sensor: Internally located at device inlet, digital mass flow type, -70 to +180 L/min |
Maximum single fault steady pressure |
Device will shut down in the presence of a single fault if the steady state pressure exceeds: 30 cm H2O (30 hPa) for more than 6 sec or 40 cm H2O (40 hPa) for more than 1 sec. |
Sound |
Pressure level measured according to EN ISO 17510-1:2009 (CPAP mode): SlimLine: 26.6 dBA with uncertainty of 2 dBA Standard: 26.6 dBA with uncertainty of 2 dBA Power level measured according to EN ISO 17510-1:2009 (CPAP mode): SlimLine: 34.6 dBA with uncertainty of 2 dBA Standard: 34.6 dBA with uncertainty of 2 dBA |
DIMENSIONS |
116 mm x 205 mm x 150 mm |
Temperature |
Maximum heater plate: 68°C Cut-out: 74°C Maximum gas temperature: £ 41°C |
Air filter |
Standard: Material: Polyester non woven fibre Average arrestance: >75% for ~7 micron dust Hypoallergenic: Material: Acrylic and polypropylene fibres in a polypropylene carrier Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron dust |
Operating pressure range APAP, CPAP |
4 to 20 cm H2O (4 to 20 hPa) |
Supplemental oxygen Maximum flow |
4 L/min |
NOTE |
The manufacturer reserves the right to change these specifications without notice. · Do not use electrically conductive or antistatic air tubing.