It makes a breathing more comfortable and natural lowering the pressure for exhalation. Heart disease prevention by treating a sleep apnea it can reduce a heart disease. It improved emo.It has heated humidification is a desirable feature in a CPAP especially in dry climate. It helps to prevent a sore throat ,dry sinus passengers or even infections. It made with the heating plate built right into the machine. It The RESMED AIRSENSE 10 CPAP is a basic fixed pressure device with integrated humidifier and cellular connectivity it with each Airsense 10 device offering built in cellular technology . It access the therapy data and remote access to device settings it helping to more connected and efficient . Auto ramp with one set detection. Easy breathe waveform with EPR. It has built in heated humidifier. It has desirable new features such as including a color LCD display with integrated light sensor a smart start features ramp and expiratory pressure features for added comfort and the option of heated tube keeps the temperature of the entire air circuit at a steady level. The machines auto climate control mode goes on step further by detecting the room temperature and the adjusting level of humidification delivered the mask. It is compact for travelling with a full sized machine. It improved eliminate snoring. It built in ambient light sensor.
• Continuous positive airway pressure (CPAP) is a method for patient who have sleep apnea disorders.obstructive sleep apnea is a sleep disorders when your breathing stop and starts.
• Obstructive sleep apnea causes interruption breathing often because your throat or airway temporarily block them.
• A CPAP machine works a compressor motor generates a continuous steam of pressurized air that travel through an air filter into a flexible tube.
• The machine works on a steady floe of pressurized air into your nose and mouth as you sleep it keep your airways helps breathe normally.
• The airstream from the CPAP machine pushes against any blockages opening your airways so your lungs receives plenty of oxygen.
• It improves your ability to sleep without walking up.
• It lowers your risk of having a heart attack stroke or other cardiovascular attack.
• It hepls lower your blood pressure.
• It reduces daytime sleepiness.
• It may reduce blood glucose and cholesterol level.
• Blood pressure
• Heart disease
• Reduce stroke risk
• Upper airway resistance syndrome
• Improves your ability to sleep without waking up
• Lowers your risk of having a heart attack, stroke, or other cardiovascular event
• Helps lower your blood pressure
• Reduces daytime sleepiness
• May reduce blood glucose and cholesterol levels
POWER SUPPLY UNIT AC input range DC output Typical power consumption Peak power consumption | 100–240V, 50–60Hz 1.0–1.5A, Class II 115V, 400Hz 1.5A, Class II (nominal for aircraft use) 24V 3.75A 53W (57VA) 104W (108VA) |
Environmental conditions | Operating temperature: +41°F to +95°F (+5°C to +35°C) Note: The air flow for breathing produced by this therapy device can be higher than the temperature of the room. Under extreme ambient temperature conditions (104°F/40°C) the device remains safe. Operating humidity: 10 to 95% relative humidity, non-condensing Operating altitude: Sea level to 8,500' (2,591 m); air pressure range 1013 hPa to 738 hPa Storage and transport temperature: -4°F to +140°F (-20°C to +60°C) Storage and transport humidity: 5 to 95% relative humidity, non-condensing |
Electromagnetic compatibility | The AirSense 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC 60601-1- 2:2014, for residential, commercial and light industry environments. It is recommended that mobile communication devices are kept at least 1 m away from the device. Information regarding the electromagnetic emissions and immunity |
Sensors | Pressure sensor: Internally located at device outlet, analog gauge pressure type, 0 to 40 cm H2O (0 to 40 hPa) Flow sensor: Internally located at device inlet, digital mass flow type, -70 to +180 L/min |
Maximum single fault steady pressure | Device will shut down in the presence of a single fault if the steady state pressure exceeds: 30 cm H2O (30 hPa) for more than 6 sec or 40 cm H2O (40 hPa) for more than 1 sec. |
Sound | Pressure level measured according to ISO 80601-2-70:2015 (CPAP mode): SlimLine: 25 dBA with uncertainty of 2 dBA Standard: 25 dBA with uncertainty of 2 dBA SlimLine or Standard and humidification: 27 dBA with uncertainty of 2 dBA Power level measured according to ISO 80601-2-70:2015 (CPAP mode): SlimLine: 33 dBA with uncertainty of 2 dBA Standard: 33 dBA with uncertainty of 2 dBA SlimLine or Standard and humidification: 35 dBA with uncertainty of 2 dBA |
DIMENSIONS | 4.57" x 10.04" x 5.91" (116 mm x 255 mm x 150 mm) |
TEMPERATURE | Maximum heater plate: 154°F (68°C) Cut-out: 165°F (74°C) Maximum gas temperature: ≤ 106°F (≤ 41°C) |
Air filter | Standard: Material: Polyester non woven fiber Average arrestance: >75% for ~7 micron dust Hypoallergenic: Material: Acrylic and polypropylene fibers in a polypropylene carrier Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron dust |
Operating pressure range AutoSet, AutoSet For Her, Elite, CPAP | 4 to 20 cm H2O |
Supplemental oxygen Maximum flow | 4 L/min |
Humidifier performance | Mask Pressure cm H2O (hPa) RH output % at 63°F (17°C) ambient temperature RH output % at 72°F (22°C) ambient temperature Nominal system output AH1 , BTPS2 Setting 4 Setting 8 Setting 4 Setting 8 4 85 100 6 >10 10 85 100 6 >10 20 85 90 6 >10 1 |
NOTE | The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should not be fitted to the mask. • Do not use electrically conductive or antistatic air tubing. • The temperature and relative humidity settings displayed are not measured values. |