It is an auto adjusting bilevel device that use comfort of AUTO SET algorithm and it easy to breathe waveform to treat the obstructive sleep apnea it benefits for greater pressure support .This device works by adjusting baseline pressure to hold the airway to maintaining a fixed pressure support . It is dependable V sync continuous leak management enables reliable delivery of therapy pressure while maintaining a patient synchrony. It can customizable of TI control to set limit of patient inspiratory time while on a spontaneous breathing . The air curve v auto records the therapy information on the machine board memory .The machine can also function as a CPAP machine . In auto mode the machine will automatically adjust the exact pressure needed. It helps to adjust IPAP (inhalation ) EPAP (exhalation ) pressure minimizing sleep disturbance . The humidifier moisture the air is designed to make therapy more comfortable it can set the humidity level to off or between 1 to 8
Categories: BILEVEL, BILEVEL / NIV/ BPAP
• A BIPAP is a Non invasive therapy it a pressure controlled in a spontaneous breathing.
• It is used to help you breathe easier when you have condition that make breathing difficult .
• Obstructive sleep apnea is a serious condition in which your breathing stop and starts repeatedly while you are asleep.
• Chronic obstructive pulmonary disease (copd)
• Asthma
• Obesity hypoventilation syndrome.
• Heart condition such as congestive heart failure or coronary artery disease
• A neurological or neuromuscular disorder that affect your ability to breathe.
• Poor breathing after a medical procedure such as surgery.
• The machine works by pushing the pressurized air into your lungs.
• It reduces burden on respiratory muscles and lungs to inhale and exhale
• The machine motor blows air through the tube certain amount of required pressure.
• The pressurized air enters your body through the mask or plugs.
• It eliminates need for using a ventilator that makes uses of a tube that goes down into your throat without used for tracheotomy.
• A tracheotomy is a procedure to create a hole in your neck to let you breathe through the trachea.
• Easy-Comfort Breathe Technology.
• Consistent Humidification.
• Built-in Ambient Light Sensor.
• Continuous Leak Management.
• Quiet Easy-Breathe Motor.
• The built-in ambient light sensor adjusts to the brightness of the room. So if you need to get up during the night, the screen's back light won't disrupt your bed partner. The light also turns off automatically, so you can settle back to sleep without disturbance.
power supply unit | AC input range: 100–240V, 50–60Hz 1.0–1.5A, Class II 115V, 400Hz 1.5A, Class II (nominal for aircraft use) DC output: 24V 3.75A Typical power consumption: 53W (57VA) Peak power consumption: 104W (108VA) |
Environmental conditions | Operating temperature: +41°F to +95°F (+5°C to +35°C) Note: The air flow for breathing produced by this therapy device can be higher than the temperature of the room. Under extreme ambient temperature conditions (104°F/40°C) the device remains safe. Operating humidity: 10 to 95% relative humidity, non-condensing Operating altitude: Sea level to 8,500' (2,591 m); air pressure range 1013 hPa to 738 hPa Storage and transport temperature: -4°F to +140°F (-20°C to +60°C) Storage and transport humidity: 5 to 95% relative humidity, non-condensing |
Electromagnetic compatibility | The AirCurve 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC 60601-1- 2:2014, for residential, commercial and light industry environments. It is recommended that mobile communication devices are kept at least 1 m away from the device |
Sensors | Pressure sensor: Internally located at device outlet, analog gauge pressure type, 0 to 40 cm H2O (0 to 40 hPa) Flow sensor: Internally located at device inlet, digital mass flow type, -70 to +180 L/min |
Maximum single fault steady pressure | Device will shut down in the presence of a single fault if the steady state pressure exceeds: 30 cm H2O (30 hPa) for more than 6 sec or 40 cm H2O (40 hPa) for more than 1 sec |
Sound | Pressure level measured according to ISO 80601-2-70:2015 (CPAP mode): SlimLine: 25 dBA with uncertainty of 2 dBA Standard: 25 dBA with uncertainty of 2 dBA SlimLine or Standard and humidification: 27 dBA with uncertainty of 2 dBA Power level measured according to ISO 80601-2-70:2015 (CPAP mode): SlimLine: 33 dBA with uncertainty of 2 dBA Standard: 33 dBA with uncertainty of 2 dBA SlimLine or Standard and humidification: 35 dBA with uncertainty of 2 dBA |
Dimensions (H x W x D) | 4.57" x 10.04" x 5.91" (116 mm x 255 mm x 150 mm) |
Temperature | Maximum heater plate: 154°F (68°C) Cut-out: 165°F (74°C) Maximum gas temperature: ≤ 106°F (≤ 41°C) |
Air filter | Standard: Material: Polyester non woven fiber Average arrestance: >75% for ~7 micron dust Hypoallergenic: Material: Acrylic and polypropylene fibers in a polypropylene carrier Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron dust |
Operating pressure range | S, ST, T: 3 to 25 cm H2O (3 to 25 hPa) VAuto: 4 to 25 cm H2O (4 to 25 hPa) CPAP: 4 to 20 cm H2O (4 to 20 hPa) |
Supplemental oxygen Maximum flow: | For VAuto device: 4 L/min (CPAP, S, VAuto) For S device: 4L/min (CPAP, S) For ST device: 15 L/min (CPAP, S, ST, T) |
Humidifier performance | Mask Pressure cm H2O (hPa) RH output % at 63°F (17°C) ambient temperature RH output % at 72°F (22°C) ambient temperature Nominal system output AH1 , BTPS2 Setting 4 Setting 8 Setting 4 Setting 8 3 85 100 6 >10 4 85 100 6 >10 10 85 100 6 >10 20 85 90 6 >10 25 85 90 6 >10 |
NOTE | The manufacturer reserves the right to change these specifications without notice. • The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should not be fitted to the mask. • Do not use electrically conductive or antistatic air tubing. • The temperature and relative humidity settings displayed are not measured values. |