• Lumis 150 VPAP ST-A is a non-invasive ventilator with a set of fixed and adjustable alarms that deliver personalized ventilation. • It has unique volume assurance mode iVAPS (intelligent Volume Assured Presssure Support). • It maintains target alveolar ventilation to suit each patients changing needs. • It is an intelligent backup rate (iBR) and optional Auto EPAP, iVAPS continuously monitors ventilation and upper airway to help you further personalise therapy for your patients.
Categories: BILEVEL, BILEVEL / NIV/ BPAP
• A BIPAP is a Non invasive therapy it a pressure controlled in a spontaneous breathing. • It is used to help you breathe easier when you have condition that make breathing difficult . • Obstructive sleep apnea is a serious condition in which your breathing stop and starts repeatedly while you are asleep. • Chronic obstructive pulmonary disease (copd) • Asthma • Obesity hypoventilation syndrome. • Heart condition such as congestive heart failure or coronary artery disease • A neurological or neuromuscular disorder that affect your ability to breathe. • Poor breathing after a medical procedure such as surgery. • The machine works by pushing the pressurized air into your lungs. • It reduces burden on respiratory muscles and lungs to inhale and exhale • The machine motor blows air through the tube certain amount of required pressure. • The pressurized air enters your body through the mask or plugs. • It eliminates need for using a ventilator that makes uses of a tube that goes down into your throat without used for tracheotomy. • A tracheotomy is a procedure to create a hole in your neck to let you breathe through the trachea.
• iVAPS Technology • intelligent Backup Rate (iBR) • AutoEPAP maintains upper airway patency • Vsync • TiControl • 5 Trigger and Cycle Senstivities • QuickNav for low touch therapy adjustments • Climate Control Auto for automatic humidification • Ramp and Rampdown for extra comfort • Supported by ResScan 5.4 • Alarms for monitoring and confidence • Lumis also features technologies designed to manage any mask leak, and stay in sync with your breathing. • Easy to use.
90W power supply unit AC input range
DC output
Typical power consumption Peak power consumption |
100–240V, 50–60Hz 1.0–1.5A, Class II 115V, 400Hz 1.5A, Class II (nominal for aircraft use) 24V 3.75A 53W (57VA) 104W (108VA) |
Environmental conditions Operating temperature Operating humidity Operating altitude Storage and transport temperature Storage and transport humidity | +5°C to +35°C 10 to 95% relative humidity, non-condensing Sea level to 8,500' (2,591 m); air pressure range 1013 hPa to 738 hPa -20°C to +60°C 5 to 95% relative humidity, non-condensing |
Electromagnetic compatibility | The Lumis complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC 60601-1-2:2014, for residential, commercial and light industry environments. It is recommended that mobile communication devices are kept at least 1 m away from the device. |
Sensors Pressure sensor
Flow sensor | Internally located at device outlet, analogue gauge pressure type, 0 to 40 cm H2O (0 to 40 hPa) Internally located at device inlet, digital mass flow type, -70 to +180 L/min |
Maximum single fault steady pressure | The device will shut down in the presence of a single fault if the steady state pressure exceeds: 30 cm H2O (30 hPa) for more than 6 sec or 40 cm H2O (40 hPa) for more than 1 sec. |
Sound Standard Standard and humidification Power level Standard Standard and humidification |
25 dBA with uncertainty of 2 dBA 27 dBA with uncertainty of 2 dBA
33 dBA with uncertainty of 2 dBA 35 dBA with uncertainty of 2 dBA |
Physical - device and humidifier Dimensions (H x W x D) Air outlet Weight Housing construction Water capacity Cleanable humidifier - material | 116 mm x 255 mm x 150 mm 22 mm 1268 g Flame retardant engineering thermoplastic To maximum fill line 380 mL Injection moulded plastic, stainless steel and silicone seal |
Temperature Maximum heater plate Cut-out Maximum gas temperature |
68°C 74°C ≤ 41°C |
Air filter Standard Hypoallergenic | Material: Polyester non-woven fibre Average arrestance: >75% for ~7-micron dust Material: Acrylic and polypropylene fibres in a polypropylene carrier Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron dust |
Operating pressure range S, ST, T, PAC, iVAPS CPAP |
2 to 25 cm H2O (2 to 25 hPa) 4 to 20 cm H2O (4 to 20 hPa)
|
Supplemental oxygen Maximum flow | 15 L/min (S, ST, T, PAC, CPAP), 4 L/min (iVAPS) |