• H-80 series Respiratory High-flow Therapy Device is for the treatment of spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases. This device is for patients by prescription in the home or hospital/institutional environment. • The device is only allowed to be used after the physician provides guidance for the patient and targeted setting of the device; the physician should ensure that the patient understands all functions and methods of operation of the device. • The HFNC oxygen therapy is carried out using air \ oxygen blender, active humidifier single heated tube and nasal cannula it is respiratory support. • The HFNC continuous to increase easily set up it may also create positive pharyngeal pressure to reduce work of breathing the device midway classical oxygen delivery system like high concentrations CPAP generators . • It is designed a heated humifided mixture of air and oxygen at a high flow than the inspiratory flow. • It adjust the flow rate and body weight provides a degree of distending pressure. It also decrease works of breathing and oxygenation . • It utilizes a soft large bore nasal prongs it is similar to used on mechanical ventilator with the HFNC. It overcome the leading causes of discomfort seen with conventional oxygen therapy. • It is a difficult quality level of PEEP generated by the high flow. It support breathing in people with acute heart failure , Acute respiratory distress syndrome. • It avoids rebreathing of co2 gas and therefore decreased anatomize dead space. It promoted a slow and deep breathing .it airway hydrated and it improved mucus clearance . • This improves FRC and reduce work of breathing it flows used are high humidification is necessary to avoid drying of respiratory secretions and for maintaining nasal cilia function . • It is mild positive airway pressure and lung volume recruitment. • It will titrate FIO2 and flow to promote comfort and oxygen saturation. It ensure the monitor spo2 saturation and pleth wave form. • It ensure maintain a o2 saturation on NRB mask • It delivery of optimally humidified and warm gases • It decreased body energy expenditure. • It is used for intubation and extubation support as well as treatment and prevention of post operative respiratory failure and for oxygenation patients during trails or bronchoscopy.
Categories: HFNC / HIGH FOW, HIGH FLOW
• The HFNC oxygen therapy is carried out using air \ oxygen blender, active humidifier single heated tube and nasal cannula it is respiratory support. • The HFNC continuous to increase easily set up it may also create positive pharyngeal pressure to reduce work of breathing the device midway classical oxygen delivery system like high concentrations CPAP generators. • It is designed a heated humifided mixture of air and oxygen at a high flow than the inspiratory flow. It adjusts the flow rate and body weight provides a degree of distending pressure. • It also decrease the work of breathing and oxygenation. It utilizes a soft large bore nasal prongs it is similar to used on mechanical ventilator with the HFNC. • It overcomes the leading causes of discomfort seen with conventional oxygen therapy. It is a difficult quality level of PEEP generated by the high flow. It support breathing in people with acute heart failure, acute respiratory distress syndrome. • It avoids rebreathing of co2 gas and therefore decreased anatomize dead space. It promoted a slow and deep breathing. It airway hydrated and it improved mucus clearance. • This improves FRC and reduce work of breathing it flows used are high humidification is necessary to avoid drying of respiratory secretions and for maintaining nasal cilia function. • It is mild positive airway pressure and lung volume recruitment. It will titrate FIO2 and flow to promote comfort and oxygen saturation. It ensures the monitor spo2 saturation and pleth wave form. It ensures maintain an o2 saturation on NRB mask It delivery of optimally humidified and warm gases It decreased body energy expenditure. It is used for intubation and extubation support as well as treatment and prevention of post operative respiratory failure and for oxygenation patients during trails or bronchoscopy.
• Advanced dual fan design, two levels of HFlow and LFlow • Provide a maximum of 80L/min air/oxygen mixed gases, oxygen concentration can be set automatically • 9-stage humidification temperature: 29℃-37℃, accurate to 1℃ • Two levels of HFlow and LFlow, 2L/min~80L/min, can meet patients of most ages • Comprehensive monitoring indexes: real-time monitoring of oxygen concentration, flow, temperature, respiratory rate, sPEEP. • Innovative High Flow model AutoFlow TM Automatically set the output flow parameters based on the peak expiratory flow(PEF) of the patient, simplifying the setting steps, satisfying the treatment effect and improving patient comfort • SmartFlow Follow the patient's every breath, switch the flow between breaths, provide a higher flow when inhaling, and a lower flow when exhaling, follow your every breath like a bi-level ventilator, comfortable and more economical oxygen resources. • Hot standby function designed for convenient use and saving oxygen
90W power supply unit
AC input range DC output
Typical power consumption Peak power consumption |
100–240V, 50–60Hz 1.0–1.5A, Class II 115V, 400Hz 1.5A, Class II (nominal for aircraft use) 24V 3.75A 53W (57VA) 104W (108VA) |
Environmental conditions Operating temperature Operating humidity Operating altitude
Storage and transport temperature Storage and transport humidity |
+5°C to +35°C 10 to 95% relative humidity, non-condensing Sea level to 2,591 m; air pressure range 1013 hPa to 738 hPa -20°C to +60°C 5 to 95% relative humidity, non-condensing |
Electromagnetic compatibility | The Lumis complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC 60601-1-2:2014, for residential, commercial and light industry environments. It is recommended that mobile communication devices are kept at least 1 m away from the device. |
Sensors Pressure sensors
Flow sensor
| Internally located at device outlet, analogue gauge pressure type, 0 to 40 cm H2O (0 to 40 hPa) Internally located at device inlet, digital mass flow type, -70 to +180 L/min |
Maximum single fault steady pressure | Device will shut down in the presence of a single fault if the steady state pressure exceeds: 30 cm H2O (30 hPa) for more than 6 sec or 40 cm H2O (40 hPa) for more than 1 sec. |
Alarm volume settings | Low (nominal 54 dBA), Medium (nominal 60 dBA), High (nominal 73 dBA) |
Physical - device and humidifier Dimensions (H x W x D) Air outlet Weight Housing construction Water capacity Cleanable humidifier - material | 116 mm x 255 mm x 150 mm 22 mm 1336 g Flame retardant engineering thermoplastic To maximum fill line 380 mL Injection moulded plastic, stainless steel and silicone seal |
Temperature Maximum heater plate Cut-out Maximum gas temperature |
68°C 74°C ≤ 41°C |
Air filter Standard Hypoallergenic | Material: Polyester non-woven fibre Average arrestance: >75% for ~7 micron dust Material: Acrylic and polypropylene fibres in a polypropylene carrier Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron dust |