BMC GII Auto Cpap

CODE: HH0089
26,400.00 40,000 34% Off

Ratings and Reviews

0.0
0 Ratings
&
0 Reviews
5
0
4
0
3
0
2
0
1
0
View All Reviews

BMC GII Auto Cpap Features.

Features

AUTO Mode (Pressure Adjusts Automatically)
Titration Mode (Pressure Adjusts Automatically)
CPAP Mode (Fixed Pressure)
Heated humidifier; OPTIONAL
Minimum Pressure 4cmH20
Maximum Pressure 20cmH20
Exhalation Pressure Relief setting
Auto airflow start
Auto airflow stop
Ramp Time 0 ~ 60 Minutes
Large 2.4 inch colour screen
On screen review of last night's sleep data
Apnea sensitivity adjustment
Mask removal alarm
Power failure alert
Sound level <30dB
SD Card (4GB)
Remote sleep data reports (using iCodes)
AC Power 110V~240V
Android App available
iPhone App available

BMC GII CPAP Specification.

Dimensions

17cm*19.6cm*11.8cm (without the humidifier)

Weight

1.5 kg (without the Humidifier)

Transport and storage

Operation Transport and Storage

Transport and Storage

Temperature

5°C to 35°C (41˚F to 95˚F)

-25°C to 70°C (-13˚F to 158˚F)

Humidity

15% to 93% Non-condensing

up to 93% Non-condensing

Atmospheric Pressure:

760 ~ 1060 hPa

760 ~ 1060 hPa

Mode of Operation

Continuous

Work Mode

For E-20C system: CPAP

For E-20A system: CPAP, Auto

SD Card

With a capacity ≥ 2 G, the SD card can record patient data and fault information. Furthermore, the language pack stored on the SD card enables you to change the language of the device.

AC Power Consumption

100 ~ 240 V AC, 50 / 60 Hz, 2.0 A max

Type of Protection Against Electric Shock

Class II Equipment

Degree of Protection Against Electric Shock

Type BF Applied Part

Degree of Protection Against Ingress of Water

IP22

Pressure Range

4 to 20 hPa (in 0.5 hPa increments), ≤ 30 hPa under single fault conditions

Pressure Display Accuracy

±(0.5 hPa+4%)

Pressure Stability

4 to 20 hPa (±1 hPa)

Ramp

The ramp time ranges from 0 to 60 minutes

Sound Pressure Level

< 30 dB, when the device is working at the pressure of 10 hPa.

Sound Power Level

< 38 dB, when the device is working at the pressure of 10 hPa.

Air Hose

Length: 6 ft. (1.83 m)

Maximum Flow

Test Pressure (hPa)

4

9

15

20

Average Flow at the Patient Connection Port (L/min)

80

92

91

96